Operating Instruction TNI soft Flow 50 homecare system
Product-
name
Producer Basis
material
Con-
centra-
tion
Plastic Elas-
to-
mere
SporcidInstr.
disinfektion)
Fresenius
Instrumentdisinfectionincl.
cleaningusedforthermo-
labileandthermostable
instruments,endoscopes.
Seemoreat:
http://www.epgonline.
org//0rcid%C2%AE/#sthash.
sobiobxQ.dpuf
Glutaral4,5
g,Formalde-
hyde7,6g.
Sekusept ECOLAP
GmbH
Cleaningactiveinstru-
mentdisinfectiontion
freeofaldehydes
Glucoprota-
min
5.1.3)Housingsurfaces
Cleaningcycle:daily
Wipethehousingsurfaceofthedevicedownwithadampcloth,moistenedwithdisinfectants.Let
thedisinfectantworkaccordingtotheinstructionsofthemanufacturerandmakesurethatany
cleaningresidueshasbeenremoved.Calcicationsshouldbeavoided.
5.1.4)Humidierhomecarecomplete
Cleaningcycle:daily
Wipethehousingsurfaceofthe“humidierhomecarecomplete”downwithadampcloth,mois-
tenedwithbiodegradabledetergents.Rinseitwellunderrunningwater.
Calcicationsshouldbeavoided.
Cleaningcycle:weekly
Allpartsof“humidierhomecarecomplete”aresuitablefordishwashercleaningandcanbe
cleanedwithatemperatureof65°C.Thisisrecommendedonceaweek.
Cleanthepartswithbiodegradabledetergents(dishwashingdetergents).
Rinseallpartswellunderrunningwaterandmakesurethatallcleaningresiduesareremoved.
5.1.5)Dustlter
Cleaningcycle:weekly
Changingcycle:3month
Openthedustlterapatthebacksideofthedevice.Slightpressureontheprotrudingap
releasesthelock.