EasyManua.ls Logo

TNI softFlow 50 - 5) Environmental Conditions

TNI softFlow 50
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
Operating Instruction TNI soft Flow 50 homecare system
• ThesensorconnectionlocatedbelowthecarryinghandleoftheTNI soft Flow 50mayonly
beusedtoconnectthe“externaltemperaturemeasuringelement”(itemno.40641018)or
othercomponentsauthorizedbytheTNImedicalAG.Donotconnectunauthorizedcom-
ponentstothissocket!
• Pleasemakesurethatthedeviceispositionedinawaythatthepowerplugcanbediscon-
nectedwithoutdiculties.
• AnSD-Cardcanbeusedtostoreinformationindependentlyfromthedevice.
1.3.5)Environmentalconditions
• TheTNI soft Flow 50mayonlybeoperatedunderadmissibleambientconditions(see
chapter7,“Technical Data”).Operatingthedeviceunderambientconditionsoutofthe
rangedenedfortheguaranteedperformanceparameterswillresultinreducedperfor-
manceparameters.Iftheambientconditionsareoutofthegivenrange,thedeviceshould
stayino-modeduetosafetyreasons.
Inordertoensureproperuse,itisimportanttolettheTNI soft Flow 50adapttotheambi-
entconditions(roomtemperature).Pleasewaitabout2-4hoursbeforestartingupthede-
vice.ThisappliestotherstusageoftheTNI soft Flow 50andtotransportingthedevice,
i.e.thetransportconditionswereoutofthegivenrangeofambientconditions.Donotuse
thedeviceinhumidandpotentiallyexplosiveroomsorcombustibleatmosphere.
• IntendedfunctioningoftheTNI soft
Flow 50maybeimpairedwhenitisoperatedright
nexttoHFelectrosurgerydevices,debrillators,X-rays,transmittedpulses,radiofrequency
interferenceordevicesdesignedforshortwavetherapy.
• DonotusetheTNI soft
Flow 50whileperformingthemeasuresmentionedaboveordur-
ingmagneticresonancetomography(MRT,NMR,NMI).
• Activemobilephonesmayonlybeplacednexttothedevicewhenaminimumdistanceof
1misbeingkept.
• Thesystemmustnotbesetupnexttoaheaterandmustnotbeexposedtodirectsunlight
sincethismayinterferewithaproperoperationofthedevice.
• Thesensormeasuringtheambienttemperatureislocatedontherightintheinsideofthe
device.Donotpointanysourceofheat(e.g.heater,...)towardsthissensor.
• Inordertoensureasucientaircirculationaroundthedevice,aminimumdistanceof25
cmtoallsidesistobekept.
• Thedevicemaynotbecovered.
• Toavoidafastaccumulationofdustintheairlter,donotplacetheTNI soft
Flow 50 near
theground.
• Thedeviceisintendedtobeusedindoorandshouldbeplacedonanevenhorizontal
surface.
• Toavoiddamage,contaminationormalfunctioning,placethedeviceoutofreachfrom
pets,pestsorchildren
• Operatingthedeviceatalowambientairtemperaturemayresultinaformationofconden-
sateintheapplicator.Keeptheboxedapplicatorawayfromdirectsunlightandstoreitina
dryplace.

Table of Contents

Other manuals for TNI softFlow 50