3-123
Set the threshold of the coloring of the following
quantitative index of IOL Screening screen.Single
AvgK: AvgK / Real AvgK threshold settings
Threshold1: Sets the lower threshold of AvgK value.
When the AvgK value is less than the set value, it is
displayed in red.
Threshold2: Sets the upper threshold of AvgK value.
When the AvgK value is more than the set value, it
is displayed in red.
CYL: CYL / Real CYL threshold settings
Threshold: Sets the size of the upper threshold of
the CYL value.
When the CYL value is more than the set value, it is
displayed in red.
HOAs: threshold setting of HOAs [4mm]
Threshold1: Sets the rst upper threshold of HOAs
value.
Threshold2: Sets the second upper threshold of
HOAs value.
When the HOAs value is more than the "Threshold1"
value and less than the "Threshold2" value, it is
displayed in yellow.
When the HOAs value is more than the "Threshold2"
value, it is displayed in red.
SA: threshold setting of the SA [6mm]
Threshold: Sets the lower threshold of the SA value.
When the SA value is less than the set value, it is
displayed in red.

Table of Contents