52
SETTING FUNCTIONS ON SETUP SCREEN
LIST OF SETUP ITEMS
Setup items are categorized into 6 large indexes.
"Initial Settings" .........................items related to the initial status after power on
"Internal Printer" ........................items related to output from the internal print
"Comm" .....................................items related to data output with the external device
"Network Settings".....................items related to output using the LAN
"Operator Settings"....................items related to Operator ID
"Special" ....................................items related to maintenance (for service engineer only)
INITIAL SETTINGS
Initial contains settings related to the initial status after power on, clearing all measurement values,
etc.
Descriptions Options Details Initial value
Buzzer sound
OFF Buzzer does not sound.
ON
ON Buzzer sounds.
Auto Shoot
OFF Default measurement mode is Auto shoot off.
ON
ON Default measurement mode is Auto Shoot.
Printer output
OFF Internal printer is disabled.
ON
ON Internal printer is active.
Show Pacho value
OFF Pacho value is not displayed.
ON
ON Pacho value is displayed.
Show Tono average
OFF Tono average value is not displayed.
ON
ON Tono average value is displayed.
Show Adj value
OFF Adjusted value is not displayed.
ON
ON Adjusted value is displayed.
Patient No. reset
OFF Patient No. is not reset upon power on.
OFF
ON Patient No. is reset upon power on.
Display of patient ID
OFF Patient ID is not displayed.
ON
ON Patient ID is displayed.
Patient ID (Mandatory)
OFF Patient ID is not required.
OFF
ON Patient ID is required.
Device ID number
1-99
Set by ten-key display
Sets the Device ID number. 1
Display of Device ID num.
OFF Device ID is not displayed.
OFF
ON Device ID is displayed.
Start time of sleep mode
OFF Power save function is not used.
10 min
1 min Power save status in 1min after last operation.
5 min Power save status in 5min after last operation.
10 min Power save status in 10min after last operation.
20 min Power save status in 20min after last operation.
30 min Power save status in 30min after last operation.
60 min Power save status in 60min after last operation.
Date/Time Set by ten-key display Sets year, month, day, time (24hrs), minute and second
Installation
date/time
Tono display Unit
mmHg Display in mmHg
mmHg
digit Display in digit
hPa Display in hPa
Torr Display in Torr
Pacho display Unit
mm Display in mm
mm
µm Display in µm
Press average Mode
Integer Display in Integer
Integer
Real Display in Real
IOP Adjustment
OFF IOP adjustment type is OFF.
OFF
ON IOP adjustment type is ON.

Table of Contents