56
MAINTENANCE
MAINTENANCE
DAILY CHECKUPS
CLEANING THE INSTRUMENT
Dust on intraocular pressure measuring window....
Blow off dust with a blower.
Fingerprints and oil spots on intraocular pressure measuring window ....
Blow off dust with a blower and wipe the surface gently with a
camera lens cleaner using clean gauze.
Dirty instrument cover......................Wipe the surface with the attached monitor cleaner or a dry
soft cloth. Never use solvents or a chemical duster.
CLEANING THE INTRAOCULAR PRESSURE MEASURING WINDOW
To secure auto alignment and correct measurement values, clean the measuring window glass
after each day's work.
Clean the glass when "CLEAN THE MEASURING WINDOW GLASS" is displayed on the control
panel screen.
1 Prepare the ethanol.
2 Using a blower, remove dust and dirt from the glass surface.
3 Moisten the applicator with ethanol.
4 Wipe the glass surface lightly with the applicator, from the center outward.
5 Use a new applicator and wipe the glass surface in a similar manner; repeat this several times.
CAUTION
To clean the measuring window glass, measuring nozzle and the
glass inside the measuring nozzle, use ethanol. Using other
chemicals may cause damage to the patient's eye during
measurement.
NOTE
Do not apply unreasonable force to the measuring nozzle while cleaning.
To avoid problems, do not leave the cotton fibers inside.
Be sure to use only the attached applicator.
NOTE
To ensure thorough removal of grease from the intraocular pressure
measuring window, be sure to replace the applicator and use a new one for
each of these repeated wiping operations.
Wiping the glass surface