73
GENERAL INFORMATION ON USAGE AND MAINTENANCE
PATIENT'S ENVIRONMENT
When the patient or inspector may touch the devices (including the connecting devices) or when the
patient or inspector may touch the person that comes into contact with the devices (including the con-
necting devices), the patient's environment is shown below.
In the patient's environment, use the device conforming to IEC60601-1. If you are compelled to use
any device not conforming to IEC60601-1, use an insulation transformer or the common protective
earth system.
Note 1: Use the device conforming to IEC60950-1.
CAUTION
Don't connect an additional power strip or an extension cord to
the system.
Don't connect the device which is not recognized as one compo-
nent of the system.
Devices applicable to the
patient's environment
Barcode reader (Note 1)
Do not use the power strip in the
patient's environment. Connect
the power supply of the device
to the commercial power supply.
Radius 1.5m
2.5m
1.5m1.5m