TROUBLESHOOTING
"TOO CLOSE" Warns that the measuring head is too close to the patient's eye.
"TOO FAR" Warns that the measuring head is too far from the patient's eye.
"Please check the
DATE/TIME"
The battery for the built-in clock has become worn out. Before using, confirm
the time and date on the SETUP menu.
If the message comes up frequently, call your service engineer.
"Perform air check?" Displayed to confirm whether to perform air check.
"Air checked.
Press OK to continue."
Displayed when normal air check operation is confirmed.
"Confirm abnormal
action(-) of air check"
Displayed when an anomaly occurred during air check. Check the measuring
nozzle for any foreign matter.
"Confirm abnormal action(+) of
air check"
Displayed when an anomaly occurred during air check. Check the measuring
nozzle for any foreign matter.
"First Octet is 1 - 223 Range" Displayed when the first octet of IP address, default gateway, primary DNS
server or secondary DNS server is set to a value out of the specified input
range.
"Value is irregular.
Input valid value"
Displayed when the octet is set to a value out of the specified input range.
Enter a value within the input range.
"The IP address is 0 - 255
Range"
Displayed when the subnet mask is set to a value off the input rule. Enter a
value within the input rule.
"At least 3 characters are
required for operator ID prefix."
Displayed when the input examiner ID prefix is less than 3 characters. Enter a
prefix with 3 characters.
"Incorrect password" Displayed when the password inputted to select a special mode is incorrect.