105
SPECIFICATIONS AND PERFORMANCE
SAFETY OF LED PRODUCT
CAUTION
Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous radiation expo-
sure.
Do not remove the enclosures. LED high-power is radiated.
Class of LED product CLASS1 LED PRODUCT (IEC60825-1:2001)
LED OUTPUT IR LED (for fundus observation)
Aperture of LED Objective lens*
Output of cornea 800µW
Wavelength 850nm
Half width 30nm
Beam divergence 0.785rad
Pulse width CW
SPLIT LED (for split lines)
Aperture of LED Objective lens*
Output of cornea 40µW
Wavelength 870nm
Half width 35nm
Beam divergence 0.015rad
Pulse width CW
ANT OBS LED (for anterior observation)
Aperture of LED Window for anterior observation*
Output of cornea 200µW
Wavelength 950nm
Half width 50nm
Beam divergence 0.06rad
Pulse width CW
ANT OBS LED (for anterior OCT)
Aperture of LED Window for anterior OCT*
Output of cornea 100µW
Wavelength 850nm
Half width 30nm
Beam divergence 0.06rad
Pulse width CW
ZEN FIX LED (for anterior segment fixation)
Aperture of LED Window for anterior fixation*
Output of cornea 0.5µW
Wavelength 568nm
Half width 30nm
Beam divergence 0.044rad
Pulse width CW

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon DRI OCT-1 Triton and is the answer not in the manual?

Topcon DRI OCT-1 Triton Specifications

General IconGeneral
BrandTopcon
ModelDRI OCT-1 Triton
CategoryMedical Equipment
LanguageEnglish

Related product manuals