35
BASIC OPERATIONS
BASIC OPERATIONS
FUNDUS TOMOGRAPHY
FUNDUS TOMOGRAPHY
Selecting the capture icon
On the capture icon selection screen, which is the initial screen, touch and select the fundus tomogram
capture icon.
CAUTION
To avoid injury of the patient, be careful not to bump the patient's eye or
nose with the instrument and external fixation target when operating the
touch display.
NOTE
Caution in photography
Under the following photography conditions, there is a bright spot on the cen-
ter of the picture.
The patient's pupil is small. (This condition includes the case of "When the
small pupil diaphragm is set to ON".)
The flash level is high.
Alignment is not adjusted properly.
To improve these conditions, carry out the following operations.
Darken the room to enhance dilation. When using mydriatic, follow the
judgment of a doctor and the instruction manual about medicine.
Adjust alignment for the position where the flare is the least visible on the
monitor screen.
Set the flash level to the lowest setting the environment will allow.
Then, take a picture.
To ensure correct imaging, adjust the table height so the patient can relax with
his/her chin placed centrally on the chin-rest.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon DRI OCT-1 Triton and is the answer not in the manual?

Topcon DRI OCT-1 Triton Specifications

General IconGeneral
BrandTopcon
ModelDRI OCT-1 Triton
CategoryMedical Equipment
LanguageEnglish

Related product manuals