3–4
4. Subtract the measured density from the alternative density to
determine the density offset to be entered into the PaveTracker
Plus software. For example, if the density as determined by a
core analysis is 150 pcf, and the density measured by
PaveTracker Plus is 146 pcf, the density offset is +4 pcf.
NOTE
For best results, repeat the above procedure to
determine the density offset at each of several in-place
locations. Then calculate the average density offset.
MANAGING DENSITY OFFSETS
Density offsets are managed using the Density Offset menu. To
access this menu, press OFFSET. The gauge displays:
Density Offset
#.## pcf
1-Enable 2-Disable
3-Change Offset
The gauge displays the current density offset on the second line. To
enable the offset, press 1. To disable the offset, press 2.
To enter a new density offset, press 3. The gauge prompts for the
density offset as shown below:
Density Offset
#.## pcf
Select (+/-)