Error
Name
Message Description Corrective Action
Err:013 Air Valve
Error
Message is displayed when the
system can't reach the
atmosphere pressure in the
chamber by opening Air Valve
for 3 minutes.
Replace air filter.
Err:014 Low
Vacuum
The message displayed if the
system can't reach the required
vacuum conditions in the
chamber for 20 minutes.
Verify that the door
is properly closed.
Perform a new
cycle.
If the problem
persists, call the
technician.
Err:015 Low Heat This message displayed if the
system can't reach the required
pressure or temperature
conditions in the chamber after
20 minutes during the
Prevacuum and Heating stages.
Check that the
autoclave is not
overloaded.
If the problem
persists, call the
technician.
Err:016
Press
Sensor
Cutout
This message and symbol are
displayed if the chamber
pressure sensor is
disconnected or out of range.
Call the technician.
Err:017
Pt2 out This message and symbol are
displayed when the Jacket
Temperature sensor Pt2 is
disconnected or out of range.
Call the technician.
Err:018
Press
Sensor
Cutout
This message and symbol are
displayed if steam generator
pressure sensor is disconnected
or out of range.
Call the technician.
Err:020 No Gen
water
This message is displayed if
the generator electrode has no
water for a specific time.
Perform a new
cycle .
If the problem
persists, call the
technician.
Err:021 Test Fail This message is displayed if
one of the test programs fails.
Check the leakage
from the door.
Dry the chamber and
try a second time.
This message is displayed if
there is lack of water in the
mineral-free water reservoir.
Pour water in the
front funnel until it
reach the full level.
This message is displayed if
the waste water reservoir is
full.
Drain the reservoir.
This message is displayed
while the autoclave is
preheating.
Wait for the
autoclave to reach
the proper operating
conditions for
staring a new cycle.