Page 27 of 56
5.5. Program 5: Wdry 121
Program 5 is recommended for sterilizing double wrapped instruments,
porous loads and Hollow A (e.g. dental hand pieces, suction pipes)
loads at temperatures of 121ºC/250ºF with 20 minutes drying stage,
which its manufacturer declares their compliance to be sterilized in the
following conditions:
The solid load shall not exceed 8kg (17.6 lb).
The porous load shall not exceed 2kg (4.4 lb).
Nominal Parameters
• Sterilization temperature: 121ºC (250ºF).
• Sterilization time: 20 minutes.
• Dry time: 20 minutes.
• Max. cycle time: 70 minutes.
• Average cycle time: 60 minutes.
Operation Sequence
• Air removal phase; three vacuum pulses as described in the diagram
below. Saturated steam is introduced into the chamber until the
sterilization temperature is reached.
• Sterilization phase; temperature and pressure are maintained
constant at the pre-set level for the sterilization time.
• Fast exhaust phase; steam is rapidly exhausted from the chamber
and drying stage is performed with a continuous vacuum pulse,
followed by vacuum break (air inlet), until pressure equalizes
atmospheric pressure.
5.6.
Cycle start