8. INSTALLATION
8.1. Placing
CAUTION:
The sterilizer must be placed on a rigid and leveled surface. The
stand must be able to withstand the load of the device and loaded
material.
1. Counter top able to support 80 kg.
2. Counter space minimum 55cmW x 65cmD (22”W x 25”D)
* (see unit dimensions)
8.1.1 Placing the Autoclave
Keep the back and the sides of the autoclave approximately
50 mm (2”) away from the wall to allow ventilation and to
facilitate the device disconnection.
If placed in a cabinet, verify that the rear of the cabinet is
open to allow ventilation.
Insufficient space for ventilation may result in an increase
of the autoclave's temperature that may damage the
instrument.
It is recommended that enough space be left around the
autoclave to give a technician access for servicing the
machine.
8.1.2 Connections to Utility Supplies
Plug the power cord into the 1 phase, 208/240Vac,
50/60Hz, 10A supply.
8.2. Lifting and carrying
CAUTION:
Before moving the autoclave, Make sure that the electric cord is
disconnected from the power, and there is no pressure in the
chamber and in the generator.
Attention! The pressure of the generator does not decrease
immediately when the equipment is turned off. Wait approx. ½ an
hour to verify that the pressure decreased to atmospheric pressure.
1. Disconnect the power supply cord.
2. Drain the water from both reservoirs.
To avoid injuries, lifting and carrying should be done with at least two
persons or by using a fork-lift or any other mechanical aid.
Do not drop the device!