10.4. Operations
Hold the opening handle in open position, while pushing the door
until it comes to close position, then release the handle. (The open
door symbol is replaced by the next load number).
Select the program
UP key: next program
DOWN key: previous program
Verify that you choose the adequate cycle.
Start the cycle by pressing the START/STOP key
10.5. Unloading
When the cycle ends the buzzer sounds continuously during 10
seconds and the message CYCLE END is displayed on the screen.
Open the autoclave's as follows:
Open the door the minimum required to let the residual steam to
escape from the chamber. Only after there is no vapor open the door
widely.
Warning
To avoid sever injuries from hot steam when opening the door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your
body over the door.
Wear heat-resistant gloves or use the tray handle to remove the load
from the autoclave
On completion of the cycle, the load shall be visual inspected to
ascertain that the load is dry, and that sterilization indicators have
made the required color change.
10.6. Stopping the process
It is possible to stop the program while the autoclave is in operating.
Pressing the START/STOP key at any stage of process, stops the
operation.
10.7. Canceling the ERROR message
At the end of an aborted process, the error number, CYCLE FAIL
and
a blinking warning symbol are displayed on the screen. Refer
to "Displayed Error and Operational Messages".
Pressing the START/STOP key cancels the displayed message
number and enables opening the door.
Warning
The load has not completed a sterilization cycle, therefore it is not
sterile. Handle it as contaminated load.
10.8. Sleep Mode
The autoclave has a sleep mode. If the autoclave has not been used for
4 hours and no key has been pressed during this period. In this mode all
outputs are turned off. To exit the "Sleep" status, press any push-
button.