Page 40 of 56
8.3. Filling the Mineral-Free Water Reservoir.
1. Open the door (1).
2. Pour water, gently, into the front funnel (2) until it reaches the
required level (3) on the side gauge. It is preferable to use a carafe
Please note that the side gauge is divided into three sections. The
bottom red section (4) indicates that the water level in the reservoir is
too low. The middle blue section (3) shows that there is sufficient
amount of water. The upper red section (2) indicates that the water
level in the reservoir is too high.
Caution!
Under no circumstance should water be filled above the top sign
level on the side gauge.
3. If the reservoir is empty, it is recommended to add water directly
into the reservoir, as follows:
♦ Remove the water reservoir cover (6).
♦ Pour distilled water into the reservoir through the opening on
top of the autoclave until it reaches the base of the safety valve
holder (7).
Caution:
Under no circumstance should water be filled above the safety valve
holder.
In case more water is added accidentally above the top sign level on the
side gauge, decrease the water level by draining the reservoir before
starting a cycle (see para. 11.2).
Caution:
Use only water having the characteristics as per table in para. 2.9.
Tap water may clog the system and using them can invalidate the
manufacturer’s guarantee.