11.6. Cleaning water strainer
CAUTION!
Before proceeding, Make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
CAUTION!
Do not touch the strainer shortly after operation.
Touching the hot strainer’s cover may cause severe injuries.
If maintenance operation is performed while strainer cover is hot,
use heat resistant gloves to avoid injuries.
1. Open the door.
2. Remove the trays and the trays holder.
3. Remove the strainer element mounted on the bottom-rear of the
chamber.
4. Rinse the strainer with water using a brush if necessary.
5. Reinstall the strainer element.
6. Reinstall the trays holder and the trays.
Strainer element