Page 48 of 56
11.2. Draining the Reservoirs
(Applies to the clean-water reservoir and to the waste-water reservoir)
1. To drain the reservoir, use item (5) with the plastic hose (6)
attached to it (supplied with the autoclave).
2. Insert part (5) into valve (3) and press it until you hear a “click”.
The drain valve is open.
3. When the water reservoir is empty, press part (4). Item (5) will pop
out approx. 3mm and the drain valve will be closed. Remove item
(5) with the plastic tube.
4. If the drained reservoir is the clean-water reservoir, fill reservoir
with distilled water until, reaches full level. (approximately 6.5
liters (1.72 gallons)).
The autoclave is now ready for use.
No. description
1 Clean water drain
2 Waste water drain
Press this direction
to o