Page 26 of 83 Pages
5.2 Program 2: Wrapped instruments
For wrapped instruments, when the manufacturer recommends autoclaving
at temperatures of 134
C (273F).
Nominal parameters default settings
Sterilization temperature: 134C (273F)
Sterilization time : 7 minutes
Operations sequence:
Insert water to chamber and heating by actuation of electrical heaters
until reaching the high preset air pulse.
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant during the sterilization
time.
Fast exhaust, steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.