Page 32 of 83 Pages
5.8 Program 8: Custom 2
Sterilization parameters are adjustable for items that cannot be sterilized in
any of the preceding default programs.
Validation of the sterility when
using this program is the responsibility of the user.
Nominal parameters default settings
Sterilization temperature: 121C (250F).
Sterilization time: Default 20 minutes.
Operations sequence:
Insert water to chamber and heating by actuation of electrical heaters
until reaching the high preset air pulse.
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant for the preset
sterilization time.
Fast exhaust, steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.

Table of Contents