Page 65 of 83 Pages
PRINTER OUTPUT DESCRIPTION
Operator:
____________
To be filled in manually by operator
Time: 12:14:47 Time sterilization cycle ended
Cycle Ended
00:24:43 101.3 099.7 Cycle finished time
E 00:23:43 107.0 107.4 The time, temperature and pressure during exhaust
E 00:22:08 134.5 311.9 The time, temperature and pressure during exhaust
CLK 2: 12:12:10:00
CLK 1: 12:12:10:00
S 00:22:07 134.5 311.6 The time, temperature and pressure during sterilization
S 00:22:06 134.5 311.6 The time, temperature and pressure during sterilization
S 00:21:06 134.6 311.0 The time, temperature and pressure during sterilization
S 00:20:06 134.5 310.1 The time, temperature and pressure during sterilization
S 00:19:06 134.8 311.1 The time, temperature and pressure during sterilization
S 00:18:06 134.5 315.8 The time, temperature and pressure during sterilization
CLK 2: 12:08:08:00
CLK 1: 12:08:08:00
H 00:18:04 134.4 315.1 The time, temperature and pressure during heating
H 00:16:35 128.9 268.4 The time, temperature and pressure during heating
H 00:13:35 116.3 180.9 The time, temperature and pressure during heating
A 00:11:04 107.0 120.9 The time, temperature and pressure during Air removal
A 00:10:22 115.1 180.4 The time, temperature and pressure during Air removal
A 00:10:00 113.7 170.5 The time, temperature and pressure during Air removal
A 00:07:00 098.7 101.5 The time, temperature and pressure during Air removal
A 00:06:45 097.6 101.4 The time, temperature and pressure during Air removal
A 00:03:45 080.2 099.4 The time, temperature and pressure during Air removal
A 00:00:45 053.7 099.4 The time, temperature and pressure during Air removal
A 00:00:04 046.5 100.0 The time, temperature and pressure during Air removal
TIME °C kPa
Ster Time: 4.0min Sterilization time for selected program
Ster Temp: 134.0°C Sterilization temperature in chamber for selected program
Cycle: Unwrapped instru Cycle name
Cycle Num: 000001 Cycle number
Version: 1.0.00.00
Software version A.B.CC.DD = 1.0.00.00
A: Door Type: Single Manual = 1
B: Vacuum Type = 0
C: Total number of Input/Output functionality that are not as default = 00
D: Total number of parameters values that are not as default = 00
Model: 5050ELVC-D Model name
Ser. Num: 000000000001 Model Serial number
Time: 11:50:05 Time sterilization cycle started
Date: 9/FEB/2010 Date sterilization cycle started
-----------------------
Time: 08:51:39 Time of turning on
Date: 9/FEB/2010 Date of turning on
POWER ON The device is turned on
Time: 00:00:00 Time of turning off
Date: 9/FEB/2010 Date of turning off
POWER OFF The device is turned off
Legend
A Air removal stage (Steam flash) E Exhaust stage
H Heating stage
S Sterilization stage CLK 1 Real Time Clock
C Cooling stage (for ELVC-D only) CLK 2 Software clock