EasyManua.ls Logo

Tuttnauer ELV-D - Page 80

Tuttnauer ELV-D
85 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 78 of 83 Pages
12.2.2 ASME-approved type safety valve
Operate the sterilization cycle according to the manual.
Allow a pressure of approximately 200 kPa (29-psi) to build up
in the chamber.
Operate the safety valve by pulling the ring of the safety valve
using a tool, i.e. screwdriver, hook etc and lift the safety valve
ring for 2 seconds. Be careful not to burn your hands.
Press the
STOP/STOP key to interrupt the operation, and
exhaust steam from chamber.
Wait until pressure goes down to zero, only then can the door be
opened.
Attention:
Use protective gloves in order not to burn your hands with the
hot steam.
Safety valve ring

Table of Contents

Related product manuals