Page 81 of 83 Pages
Message /
Symbol /
Problem
Failure Description Corrective Action
Cycle Failed
This message and symbol are
displayed if an error occurs before
sterilization cycle is completed.
Perform a new cycle.
Air Error
This message is displayed at the end
of the cycle If the autoclave does not
reach the atmospheric pressure after
10 minutes.
Wait until the autoclave
reaches the atmospheric
pressure and perform a new
cycle.
Compressed
air supply
error
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the compressed
air supply.
Periodical
check time
exceeded -
Please call for
service
The periodical maintenance time has
passed.
Call for service.
Cycle counter
exceeded -
Please call for
service
Number of cycles, since last
periodical maintenance, exceeded
the preset number as defined by
"cycle counter" parameter.
Call for service.
Power Down
This message is displayed if power
down has occurred during the cycle.
(this message will print out in the
printer after the autoclave will turn
on).
Turn on the autoclave and
wait until the autoclave is
ready (reaches the safe
condition) and perform a new
cycle.
Supply
distilled water
error (digitat
input option)
This message is displayed in case of
a mineral free water supply
malfunction.
Check and fix the mineral
free water supply
Supply water
error (digitat
input option)
This message is displayed in case of
a city (tap) water supply
malfunction.
Check and fix the city (tap)
water supply.
Compressed
air supply
error (digitat
input option)
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the air supply.
No Water
This message is displayed if the
electrode in the chamber did not
sense water within the preset time.
1. check and fix the mineral
free water supply.
2.check and clean the water
inlet filter.
3. Clean the water level
electrode.