Chapter 10 __________________________________________________________ Troubleshooting
VAISALA______________________________________________________________________ 171
See also Chapter 10 Troubleshooting, on page 167.
Rain and Wind Sensor Heating
Control
The supervisor message aSU shows you continuously monitored
information about rain and wind sensor heating (heating temperature Th
and heating voltage Vh). See Supervisor Message on page 144.
Table 28 Error Messaging/Text Messages
Text message
identifier (in
NMEA 0183 v3.0
protocol only)
Text Message Interpretation and action
01 Unable to measure error The requested parameters are not activated in the
message and check the parameter section fields. See
Chapter 8 Sensor and Data Message Settings, on page
127.
02 Sync/address error The device address in the beginning of the command is
invalid. Ask the device address with the ?! (SDI-12) or
?<cr><lf> (ASCII and NMEA) command and retype the
command with the correct address.
03 Unknown cmd error The command is not supported, use the correct
command format. See Chapter 7 Retrieving Data
Messages, on page 83.
04 Profile reset Checksum error in configuration settings during power-
up. Factory settings used instead.
05 Factory reset Checksum error in calibration settings during power-up.
Factory settings used instead.
06 Version reset New software version in use.
07 Start-up Software reset. Program starts from the beginning.
08 Use chksum xxx Given checksum not correct for the command. Use the
proposed checksum.
09 Measurement reset The ongoing measurement of all the sensors interrupted
and started from the beginning.
10 Rain reset The precipitation sensor counter reset.
11 Inty reset Precipitation sensor intensity counter reset.
NOTE
When a WXT530 Series transmitter is used on an RS-485 bus with
other polled devices, the error messaging feature must be disabled
using the command: 0SU,S=N<crlf>.