User’s Guide ______________________________________________________________________
78 __________________________________________________________________ M211840EN-C
Communication Setting Commands
Checking Current Communication
Settings (aXU)
Use this command to request the current communication settings.
Command format in ASCII and NMEA 0183: aXU<cr><lf>
Command format in SDI-12: aXXU!
Example response in ASCII and NMEA 0183:
aXU,A=a,M=[M],T=[T],C=[C],I=[I],B=[B],D=[D],P=[P],S=[S],
L=[L],N=[N],V=[V]<cr><lf>
Example response in SDI-12:
aXXU,A=a,M=[M],T=[T],C=[C],I=[I],B=[B],D=[D],P=[P],S=[S],
L=[L],N=[N],V=[V]<cr><lf>
NOTE
In this section, the commands to be typed by the user are presented in
normal text while the responses of the transmitter are presented in
italic.
where
a = Device address, which can consist of the following
characters: 0 (default) ... 9, A ... Z, a ... z.
XU = Device settings command in ASCII and NMEA 0183
XXU = Device settings command in SDI-12
<cr><lf> = Command terminator in ASCII and NMEA 0183
! = Command terminator in SDI-12