Chapter 6 ________________________________________________________  Connection Options
VAISALA_______________________________________________________________________ 81
Changing the Communication 
Settings (aXU)
Use this command to change communication settings. For details on the 
available values, see the examples below and Setting Fields on page 79. 
Command format in ASCII and NMEA 0183: 
aXU,A=x,M=x,C=x,I=x,B=x,D=x,P=x,S=x,L=x<cr><lf>
Command format in SDI-12: 
aXXU,A=x,M=x,C=x,I=x,B=x,D=x,P=x,S=x,L=x!
where
A, M, C, I, 
B, D, P, S,L
= The communication setting fields, see Setting Fields 
on page 79.
x = Input value for the setting
<cr><lf> = Command terminator in ASCII and NMEA 0183
! = Command terminator in SDI-12
NOTE
When changing the serial interface and communication protocol, note 
the following:
Each serial interface requires its specific wiring and/or jumper settings 
described in Chapter 5 Wiring and Power Management, on page 53.
Change first the serial interface field C and then the communication 
protocol field M.
Changing the serial interface to SDI-12 (C=1) automatically changes 
the baud settings to 1200, 7, E, 1 and the communication protocol to 
SDI-12 (M=S).
NOTE
Reset the transmitter to validate the changes of communication 
parameters by disconnecting the service cable or using the Reset 
(aXZ) command. See Reset (aXZ) on page 84.