Vaisala WXT532 - Figure 47 Wind Measurement Averaging Method

Vaisala WXT532
235 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User’s Guide ______________________________________________________________________
202 _________________________________________________________________ M211840EN-C
0505-215
Figure 47 Wind Measurement Averaging Method
I I
AA
time
1sec
time
1sec
...
...
I > A, all comm unication p rotocols other than SDI-12 (aXU,M=S). In this example I=5 sec and A=3 sec.
I < A, all comm unication p rotocols other than SDI-12 (aXU,M=S). In this example I=2 sec and A=5 sec.
I
I
II
...
...
A
A
A
A
A
time
1sec
...
Comm unication protocol SDI-12 (aXU,M=S). In this example A=3 sec. I does not have any function in this protocol.
Issuing
measurement
command
Data
ready
A
Issuing
measurement
command
Data
ready
...
Case 1
Case 2
Case 3