Chapter 11 ____________________________________________________ Technical Specifications
VAISALA______________________________________________________________________ 177
- When the analog output option is applied, digital communication is not available.
General Conditions
- Due to the measurement frequency used in the sonic transducers, RF interference in the
200 ... 400 kHz range can disturb wind measurement.
- Any temporary element or object, such as snow, ice, or a bird, blocking the measurement path
between the ultrasonic transducer heads can affect the wind measurement accuracy or even
invalidate the output data.
- Extreme operating conditions can alter sensor performance temporarily.
Table 36 Analog mA Output Options
Parameter Description/Value
Wind speed 0 ... 20 mA or 4 ... 20 mA
Wind direction 0 ... 20 mA or 4 ... 20 mA
Load impedance 200 Ω max
Update interval 4 Hz max
Table 37 General Conditions
Property Description/Value
Housing protection class IP65 (without mounting kit)
IP66 (with mounting kit)
Temperature
operative
storage
-52...+60°C (-60...+140°F)
-60 ... +70 °C (-76 ... +158 °F)
Relative humidity 0 ... 100 %RH
Pressure 600 ... 1100 hPa
Wind 0 ... 60 m/s
CAUTION
Make sure that you power the sensor after installation. Storing the
sensor outdoors without a proper package or not powering up after
installation can affect the sensor’s expected lifespan.
NOTE
Select a heated sensor model if you operate the sensor in humid
conditions.