EasyManua.ls Logo

Vaisala WXT532 - Composite Data Message Definition (aR0)

Vaisala WXT532
235 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 8 ____________________________________________ Sensor and Data Message Settings
VAISALA______________________________________________________________________ 145
Setting Fields
[R] = Parameter selection: This field consists of 16 bits
defining the supervisor parameters included in the
data messages. The bit value 0 disables and the bit
value 1 enables the parameter.
Bits 1-8 determine the
parameters included in the
message obtained with the
following commands:
-ASCII: aR5 and ar5
-NMEA 0183: $--WIQ,XDR*hh
-SDI-12: aM5, aMC5, aC5, and
aCC5
-SDI-12 continuous: aR5 and
aRC5
1st bit (most left) Th Heating temperature
2nd bit Vh Heating voltage
3rd bit Vs Supply voltage
4th bit Vr 3.5 V reference voltage
5th bit Id Information field
6th bit spare
7th bit spare
8th bit spare
& delimiter
Bits 9-16 determine the
supervisor parameters
included in the composite data
message obtained with the
following commands:
-ASCII: aR0 and ar0
-NMEA 0183: aR0, ar0
-SDI-12: aM, aMC, aC, and aCC
-SDI-12 continuous: aR and
aRC
9th bit Th Heating temperature
10th bit Vh Heating voltage
11th bit Vs Supply voltage
12th bit Vr 3.5 V reference voltage
13th bit Id Information field
14th bit spare
15th bit spare
16th bit (most right) spare
[I] = Update interval: 1 ... 3600 seconds. When the
heating is enabled the update interval is forced to 15
seconds.
[S] = Error messaging: Y = enabled, N = disabled
[H] = Heating control enable: Y = enabled, N = disabled
Heating enabled: The control between full and half
heating power is on as described in Heating on page
34.
Heating disabled: Heating is off in all conditions.
<cr><lf> = Response terminator

Table of Contents