Appendix H __________________________________________________ Configuration Parameters
VAISALA______________________________________________________________________ 231
Table 55 Analog mA Output Parameters
Parameter Factory Range Info Command
Auxiliary
sensor
definitions
Order
option
defined
1 = Tr pt1000 temperature
1 = Ra Aux. rain amount
1 = Sl Snow level
1 = Sr Solar radiation
1 = Rt pt1000 resistance
0 = Reserved
1 = Analog output mode
0 = Reserved
[& = Delimiter]
1 = Tr pt1000 temperature
1 = Ra Aux. rain amount
1 = Sl Snow level
1 = Sr Solar radiation
1 = Rt pt1000 resistance
0 = Reserved
0 = Reserved
0 = Reserved
Format:
0000001000000000
1 = mA output is in use
0 = mA output is off
Note that you write the
parameter format
0000001000000000
but the unit answers
00000010&00000000
Note that the analog output
mode is used only in
WXT532 with the mA output
option.
IU,R
Wind speed
gain
Order
option
defined
0.000000001 ... 1000000 4 ... 20 mA = 0.266667
(20 mA = 60 m/s)
0..20 mA = 0.333333
(20 mA = 60 m/s)
SU,a
Wind speed
offset
Order
option
defined
0 ... 24 4 ... 20 mA = 4
0 ... 20 mA = 0
SU,b
Wind speed
minimum
0 0 ... 24 4 ... 20 mA = 0
0 ... 20 mA = 0
SU,c
Wind speed
maximum
20 0 ... 24 4 ... 20 mA = 20
0 ... 20 mA = 20
SU,d
Wind speed
error
indication
22 0 ... 24 4 ... 20 mA = 22
0 ... 20 mA = 22
SU,e
Wind
direction
gain
Order
option
defined
0.000000001 ... 1000000 4 ... 20 mA = 0.044444
0 ... 20 mA = 0.055556
SU,f
Wind
direction
offset
Order
option
defined
0 ... 24 4 ... 20 mA = 4
0 ... 20 mA = 0
SU,g
Wind
direction
minimum
0 0 ... 24 4 ... 20 mA = 0
0 ... 20 mA = 0
SU,h
Wind
direction
maximum
20 0 ... 24 4 ... 20 mA = 20
0 ... 20 mA = 20
SU,j
Wind
direction
error
indication
22 0 ... 24 4 ... 20 mA = 22
0 ... 20 mA = 22
SU,k