2. Choosing an Installation Site
Smart Plus (PHT-35LHS) Installation Manual 21
2.7 Installing the Warning Lamp and Door Interlock Switch
Refer to Appendix A for a complete installation guide.
This system can be equipped with a warning lamp and the Door Interlock Switch
which are activated when the X-ray is energized.
The warning lamp and Door Interlock Switch are not included with the equipment.
The warning lamp and the Door Interlock Switch must be installed by a qualified
technician.
2.8 Installing the Emergency Stop Switch
Refer to Appendix B for a complete installation guide.
Install the Emergency Stop Switch along with the main power cable in the central
distribution panel.
Install this switch so that it is within easy reach of the operator but cannot be
accidentally pressed.
The switch must be a fool-proof model.