Remote Commands
SCPI Commands Command Tables
MAP-200 Optical Switch mOSW-C1 Installation and User Guide Standard
June 2016 Document Number 22112369-120 R000 Page 65
*SAV
Description
The *SAV command stores current status of system and all the applications with specified
valid number. The status can be recalled by command *RCL <number>. The maximum of
entries for user settings is 5. If user enters a number that has been set before, the settings
with the old number will be overwritten.
Input format
*SAV <number>
Parameter
Type
Value
<number>*
Numeric
2 – 6
* <number> 0 is reserved for the factory default settings; <number> 1 is reserved for the power-off status; Those two numbers
can’t be accessed by this command.
Output format
None.
Default
0
Errors
Standard error messages.
Example
*SAV 2
Terminal input:
*SAV 2
This example saves current system and application status as number “2”.