EasyManua.ls Logo

Weinmann MEDUMAT Standard2 - Page 116

Weinmann MEDUMAT Standard2
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
116 EN MEDUMAT Standard
2
WM 68071 05/2021
5 User menu
In this menu you can set the scale for the time axis of the etCO
2
trend. The following setting options exist: auto, 5 min, 10 min,
30 min, 60 min, 120 min. At a time setting of 5 minutes or
10 minutes, the device records a determined value every
15 seconds. At a setting of 30, 60 or 120 minutes, an average
value is recorded every 30 seconds.
In the "auto" setting, the x axis is scaled automatically depending
on the duration of the application.
The unit of the etCO
2
trend is based on the CO
2
unit. This can
be selected in the operator menu under the menu item Device
configuration (see "6.3.7 Device configuration", page 129).
5.3.7 Bluetooth
®
(only with Bluetooth
®
data
transmission option)
5-8 Submenu Bluetooth
®
If the Bluetooth
®
data transmission option is unlocked and
activated in the operator menu (see "4.15 Unlocking options",
page 99), you can activate and deactivate the Bluetooth
®
connection here. If the Bluetooth
®
connection is activated, an
operation documentation system can connect to the device to
retrieve session data.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Related product manuals