5 User menu
In this menu you can set the scale for the time axis of the etCO
2
trend. The following setting options exist: auto, 5 min, 10 min,
30 min, 60 min, 120 min. At a time setting of 5 minutes or
10 minutes, the device records a determined value every
15 seconds. At a setting of 30, 60 or 120 minutes, an average
value is recorded every 30 seconds.
In the "auto" setting, the x axis is scaled automatically depending
on the duration of the application.
The unit of the etCO
2
trend is based on the CO
2
unit. This can
be selected in the operator menu under the menu item Device
configuration (see "6.3.7 Device configuration", page 129).
5.3.7 Bluetooth
®
(only with Bluetooth
®
data
transmission option)
5-8 Submenu Bluetooth
®
If the Bluetooth
®
data transmission option is unlocked and
activated in the operator menu (see "4.15 Unlocking options",
page 99), you can activate and deactivate the Bluetooth
®
connection here. If the Bluetooth
®
connection is activated, an
operation documentation system can connect to the device to
retrieve session data.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals