MEDUMAT Standard
2
EN 117
5 User menu
If you have unlocked the Bluetooth
®
connection function in
the operator menu (see "6.3.7 Device configuration", page 129),
the MAC address of the device will also be displayed in this menu.
You can then pair MEDUMAT Standard
2
with an external data
documentation system during the application.
5.3.8 NVG (Night Vision Goggles)
5-9 Submenu NVG
You can activate the NVG mode here. When the NVG mode is
activated, the device behaves as follows:
Alarm light deactivated
Audio alarm output for all alarms permanently deactivated
Line power and battery status indicators deactivated
Risk of injury from deactivated alarm light, deactivated audio
alarm output and darkened display in NVG mode!
The alarms are barely perceptible as a result of the deactivated
alarm light, the deactivated audio alarm output and the darkened
display in NVG mode. This can injure the patient.
Always monitor patients during ventilation.
Only use the NVG option in the military sector.