c. Put the monitor, enclosed in a plastic bag with a packing list, into the original shipping
carton with the original packing materials or into another appropriate shipping carton.
d. Write the Welch Allyn RMA number with the Welch Allyn address on the outside of the
shipping carton.
Recommended service intervals
To confirm that the monitor is functioning within the design specifications, perform periodic
service using the Welch Allyn Service Tool, Gold edition, as indicated in the following table.
Component Service interval Service procedure
Monitor Annually Functional verification
NIBP module Annually Functional verification and calibration
if necessary
SpO2 module Annually Functional verification
SureTemp Plus Semi-annually Functional verification
Braun ThermoScan PRO
4000
Annually Functional verification
Printer module Annually Functional verification
Battery 300 charge cycles Replace battery
Perform a complete functional verification of the monitor whenever any of the following conditions
exist:
• The monitor has been dropped or otherwise damaged
• The monitor is malfunctioning
• The case has been opened
• A part has been replaced
For details on performing the functional verification, see the section on functional verification.
Maintenance
For monitor maintenance information, see “Maintenance and service” in the Welch Allyn VSM
6000 Series Directions for Use (103501). Covered topics include the following:
• Replacing the printer paper
• Inspecting and cleaning the monitor and accessories
• Changing the battery
Configuration options
The monitor is available in multiple configurations:
• 6300: Basic. Does not include nurse call. Includes Ethernet and USB connectivity.
• 6400: Standard. Includes nurse call, Ethernet, and USB connectivity. The radio is optional.
Service manual Overview 11