EasyManua.ls Logo

Welch Allyn 6000 Series - Required Tools and Equipment; Power down the Monitor; Remove the Battery

Welch Allyn 6000 Series
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Caution Observe screw torque specifications, especially with screws that secure
directly into plastic standoffs.
Caution To avoid mismatching screws and holes, keep the screws for each piece
with that piece as you remove modules and circuit assemblies.
Required tools and equipment
#1 Phillips bit
#2 Phillips bit
#10 Torx bit
Torque driver calibrated for 6.0 in-lb ±1.0 in-lb
Torque driver calibrated for 7.5 in-lb ±0.5 in-lb
Slotted screwdriver
Tweezers
Needle-nose pliers
Spudger
Tie-wrap tool calibrated for torque specification 5 INT
Tie-wrap cutter
Soft lens wipes
Scissors or other cutting device
Canned air
A coin to open the battery door. Select a size that comfortably fits the slot.
Note For a list of the equipment required to perform functional verification and
calibration, see “Required equipment” in “Functional verification tests.”
Power down the monitor
1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.
This power-down method, which places the monitor into Standby mode, ensures that patient
measurements are retained in the monitor memory for a maximum of 24 hours. (These saved
measurements are available for recall, printing, or to send electronically to the network.) This
method also ensures that any configuration settings you have changed and saved will be
maintained at the next startup.
Note Because power is still available to charge the battery and power the monitor, the
monitor is in Standby mode.
Remove the battery
Before removing the battery, power down the monitor as described in "Power down the monitor."
1. Turn the monitor upside down to access the battery cover.
2. Locate the battery cover, indicated by .
3. Insert a coin into the slot and push to open. Choose a coin that fits comfortably into the slot.
48
Disassembly and repair Welch Allyn Vital Signs Monitor 6000 Series™

Table of Contents

Related product manuals