Patient data management messages
Status message Possible cause Suggested action
Alarm
Maximum number of patient
records saved. Oldest record
overwritten.
The maximum number of patient records has
been exceeded
Go to the Review tab and
delete old records to prevent
the alarm from appearing
when new records are saved.
Unable to access patient
information.
An error occurred when reading the patient
list or patient record during startup
Power down and restart the
monitor. If the error persists,
call for service.
Information
No data to save. No patient data is available Take or enter vital signs
before saving.
Patient ID required to save
data.
The configuration requires a patient ID to
save data
Disable Require patient
ID to save readings on
the Patient IDs tab,
available from the Data
management tab in
Advanced settings.
Clinician ID required to save
data.
The configuration requires a clinician ID to
save data
Disable Require clinician
ID to save readings on
the Clinician IDs tab,
available from the Data
management tab in
Advanced settings.
Patient ID required to send
data.
The configuration requires a patient ID to
send data
Add a patient ID.
No matching patient ID found. The patient ID does not match Try again with a valid patient
ID.
Patient list is full. Delete some
patients to add more.
The maximum number of patients was
exceeded
Delete a patient from the list
to add a new patient.
Stop intervals to select new
patient.
The monitor is set to take interval readings Stop intervals before
changing the patient.
46
Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series™