Reassembly notes
Before installing the temperature module into the monitor housing, ensure that the probe-well
seal is installed with the tab aligned with the slot or notch in the probe-well housing.
Route the flex cable up and over the temperature module when plugging it into the J6 ZIF
connector on the main board.
When replacing the temperature module or adding the option
1. When connecting the USB cable, do the following:
a. Connect the USB cable to the temperature module.
b. Position the retaining clip on the USB cable against the shoulder of the mini-B connector
with the adhesive pad facing away from the temperature module and toward the bottom.
c. Remove the backing to expose the adhesive on the retaining clip.
d. Rotate the clip on the USB cable to stick to the temperature module housing.
When adding a new temperature module
1. Follow the instructions to remove the empty temperature housing, main board, and LCD
display.
2. Replace the front temperature blanking panel (44) with the temperature mounting cover (56).
3. Install the thermal shield on the front housing, aligning the notch on the shield with the
housing.
Service manual Disassembly and repair 73