section to remove the empty housing or, in the case of the printer and radio access, the empty
space.
Option install procedures
This table lists the procedures required to install all available internal options. For step-by-step
instructions, see “Disassembly and repair.”
Caution Before installing an option, disconnect patients and power down the
monitor.
SpO2
SureTemp
Plus Printer Radio Procedure
Remove the battery.
Remove the rear housing.
Remove the communications board.
Remove the power supply cover.
Remove the SpO2 module or blanking panel.
Remove the NIBP module.
Remove the bottom housing.
Remove the top housing.
Remove the empty temperature housing.
Remove the temperature blanking panel.
Remove the printer door blank.
Remove the main board
Remove the LCD display
Install the communications board and radio.
Install the radio antenna.
Re-install the power-supply cover.
Install the temperature cover.
Install the temperature module.
Service manual Options, upgrades, and licenses 91