Malfunction Cause Remedy
Analysis gas flow
alarm
Analysis gas flow is not suffi-
cient or not available
Check setting of ball valve
(47)
check supply line
check pressure regulator
Calibration unsuccess-
ful interrupted
No calibration gas connected
to the analyser
Inlet pressure too low
Connect proper calibration
gas for “Offset Calibration”
respective “Gain Calibration”
to the analyser
Adjust inlet pressures
(please refer to Technical Da-
ta)
Special message on
display
System error
See Section 7.6.
Please note !
In the case of other malfunctions contact us or an authorised service centre to have the unit
checked. The gas analyser may only be returned into operation after it has been checked
and repaired by us or an authorised service centre.