EasyManua.ls Logo

Witt MFA 9000 - Shut-Down; Malfunctions and Rectification; Malfunction Table

Witt MFA 9000
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 15 of 46
WITT-GASETECHNIK GmbH & Co KG
Salinger Feld 4-8
: +49 (0)2302 89010
www.wittgas.com
Gas Safety- & Control Equipment
D-58454 Witten
Fax: +49 (0)2302 89013
witt@wittgas.com
6.4 Shut-down
The gas analyser has been designed for continuous operation. After a shut down of the gas analys-
ing system, the complete start-up procedure needs to be performed before full operability of the
system can be re-established!
Close all gas supplies to the gas analyser.
Disconnect the analyser from the mains.
Caution !
For technical reasons some residual gas pressure will remain trapped in the gas lines which
will be released when the inlet connections are unscrewed.
6.5 Malfunctions and rectification
6.5.1 Malfunction table
Malfunction Cause Remedy
The result of the anal-
ysis does not match
with the expected val-
ues
The concentration is not as it
was expected
Wrong measuring gas / carri-
er gas combination in MGM
selected
Analyser not correctly cali-
brated
Calibration gas settings at the
analyser module does not
match with the calibration gas
concentrations
Inlet filter clogged
Instrument internally contam-
inated by liquid
Change composition of the
gas mixture
Select adequate measuring
gas / carrier gas combination
for the analysis gas
Calibrate analyser
(Section 6.3 and 7.3)
Adjust the setting for calibra-
tion gas concentrations
(Section 7.3.1 and 7.3.2)
Replace filter
Contact
WITT-GASETECHNIK,
have analyser checked
After switching power
on, the display is not
illuminated
Instrument is not connected
to main power supply
Defective fuse
Check connection to mains
and rectify, as applicable
Check and replace fuse if
necessary. If the fuse is de-
fective immediately after re-
placing contact
WITT-GASETECHNIK,
Min. concentration
alarm
Concentration falls below
lower threshold limit of the
respective gas
Check adjustment of gas
mixer
Check gas supply
Check threshold limit
Max. concentration
alarm
Upper threshold limit of the
respective gas is exceeded
Check adjustment of gas
mixer
Check gas supply
Check threshold limit

Table of Contents

Related product manuals