7.6 System Errors
EEPROM ERROR
Error reading or writing data to or from internal FLASH-EEPROM
CAL GAIN ER Calibration gain exceeding max. allowed range
CAL OFFS ER Calibration offset exceeding max. allowed range
CAL DEV ER
Calibration deviation exceeding max. allowed range
CAL VAR ER
Calibration variation exceeding max. allowed range
BT MIN ER
Block temperature below specified range
BT MAX ER
Block temperature above specified range
BU MIN ER Bridge voltage below specified range
BU MAX ER
Bridge voltage above specified range
TC MIN ER
WLD-signal below specified range
TC MAX ER
WLD-signal above specified range
EXTERNAL ER
Error routed from input “DIN” (GND = o.k., +24V = error)