Multi 9310 pH value
ba75937e03 12/2011 31
Continuing with two-
point calibration
4. Thoroughly rinse the sensor with deionized water.
5. Immerse the sensor in the first buffer solution.
6. For measurements without temperature sensor
(e.g. when using an IDS adapter
):
Measure the temperature of the buffer manually and enter it with
<S
SS
S><T
TT
T>.
7. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
8. Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
9. If necessary, finish the calibration procedure as a single-point calibra-
tion with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.
01.07.2011
08:00
01.07.2011
08:00
10. Thoroughly rinse the sensor with deionized water.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 9310 and is the answer not in the manual?

wtw Multi 9310 Specifications

General IconGeneral
Brandwtw
ModelMulti 9310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals