Multi 9310 What to do if...
ba75937e03 12/2011 89
No stable measured
value
Obviously incorrect
measured values
14.2 Dissolved oxygen
Error message,
OFL, UFL
The measured value is outside the measuring range.
Cause Remedy
IDS pH sensor:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if neces-
sary
Temperature not stable Adjust temperature if necessary
IDS pH sensor + test sample:
Conductivity too low Use suitable IDS pH sensor
Temperature too high Use suitable IDS pH sensor
Organic liquids Use suitable IDS pH sensor
Cause Remedy
IDS pH sensor:
IDS pH sensor unsuitable Use suitable IDS sensor
Temperature difference
between buffer and test sample
too great
Adjust temperature of buffer or sam-
ple solutions
Measurement procedure not
suitable
Follow special procedure
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.