What to do if... Multi 9310
90 ba75937e03 12/2011
Error message,
Error
14.3 Conductivity
Error message,
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
14.4 General information
Sensor symbol
flashes
Cause Remedy
– Measured value outside the
measuring range
– Use a suitable IDS D.O. sensor
Cause Remedy
– Sensor contaminated – Clean the sensor
– Measured temperature value
outside the operating conditions
(display of OFL/UFL instead of a
temperature value)
– Keep to the temperature range for
the test sample
– Defective sensor – Replace sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
– Measured value outside the
measuring range
– Use suitable IDS conductivity sensor
Cause Remedy
– Sensor contaminated – Clean the sensor and replace it if
necessary
– Calibration solution not suitable – Check the calibration solutions
Cause Remedy
– Calibration interval expired – Recalibrate the measuring system