What to do if... Multi 9310
90 ba75937e03 12/2011
Error message,
Error
14.3 Conductivity
Error message,
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
14.4 General information
Sensor symbol
flashes
Cause Remedy
Measured value outside the
measuring range
Use a suitable IDS D.O. sensor
Cause Remedy
Sensor contaminated Clean the sensor
Measured temperature value
outside the operating conditions
(display of OFL/UFL instead of a
temperature value)
Keep to the temperature range for
the test sample
Defective sensor Replace sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Measured value outside the
measuring range
Use suitable IDS conductivity sensor
Cause Remedy
Sensor contaminated Clean the sensor and replace it if
necessary
Calibration solution not suitable Check the calibration solutions
Cause Remedy
Calibration interval expired Recalibrate the measuring system