Settings Multi 9310
72 ba75937e03 12/2011
9.5.5 Display illumination
The meter automatically switches off the display illumination if no key is
pressed for 20 seconds.
The illumination is switched on with the next keystroke again.
You can also generally switch on the display illumination (see section 9.5.1
S
YSTEM, page 70).
9.6 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1 Resetting the measurement settings
pH The following settings for pH measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting Default settings
Buffer AutoCalTEC
Calibration interval 7 d
Unit for slope mV/pH
Measured parameter pH
Unit for slope 0.001
mV resolution 0.1
Asymmetry 0 mV
Slope -59.2 mV
Man. temperature 25 °C
Single-point calibration off