Multi 9310 Conductivity
ba75937e03 12/2011 57
ting for automatic Stability control (see section 9.5.3 AUTOMATIC STABILITY CON-
TROL, page 71) in the System menu.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
8.1.2 Measuring the temperature
For reproducible conductivity measurements, it is essential to measure the
1. Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.
2. Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the display
of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3. Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.
Measured parameter Time interval Stability in the time interval
Conductivity
ϰ 10 seconds ϰ: better than
0.5% of measured value
Temperature 10 seconds
T (° C): Better than 0.02