Conductivity Multi 9310
58 ba75937e03 12/2011
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.
8.2 Temperature compensation
The calculation of the temperature compensation is based on the preset refer-
ence temperature, 20 °C or 25 °C. It appears on the display as Tr20 or Tr25.
You can select one of the following temperature compensation methods:
z Nonlinear temperature compensation (nLF) according to EN 27 888
z Linear temperature compensation (Lin) with adjustable coefficients of
0.000 ... 3.000 %
/K
z
zz
z No temperature compensation (off)
Application tips Select the following temperature compensations given in the table according to
the respective test sample:
The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 9.4.1 S
ET-
TINGS FOR IDS CONDUCTIVITY SENSORS, page 68).
Test sample Temperature compensation Display
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water nLF
according to EN 27 888
nLF
Other aqueous solu-
tions
Lin
Set linear temperature coefficient
0.000 ... 10.000 %/K
Lin
Salinity (seawater) Automatic nLF according to IOT
(International Oceanographic
Tables)
Sal, nLF