EasyManuals Logo

wtw Multi 9310 User Manual

wtw Multi 9310
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
Multi 9310 pH value
ba75937e03 12/2011 33
5.2.4 Carrying out manual calibration (ConCal)
Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is
correctly selected (see section 9.1.1 S
ETTINGS FOR PH MEASUREMENTS, page
62).
Use any one to five buffer solutions in ascending or descending order.
20. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
21. If necessary, use <M> to finish the calibration.
The calibration record is displayed.
or
Switch to calibration with the next buffer with <ENTER>.
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1. Connect the pH sensor to the meter.
The pH measuring window is displayed.
2. Keep the buffer solutions ready.
3. Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
01.07.2011
08:00

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 9310 and is the answer not in the manual?

wtw Multi 9310 Specifications

General IconGeneral
Brandwtw
ModelMulti 9310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals