pH value Multi 9310
38 ba75937e03 12/2011
5.2.6 Calibration data
Displaying the
calibration data
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL__> key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <ENTER> key in the measured value display.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item Setting/func-
tion
Description
Calibration /
Calibration data stor-
age /Display
- Displays the calibration records.
Further options:
z Scroll through the calibration
records with <S
SS
S><T
TT
T>.
z Output the displayed calibration
record to the interface with
<F2>/[USB output].
z Output all calibration records to
the interface with <F2__>[USB
output].
z Quit the display with <F1>/
[Back] or <ENTER>.
z Switch directly to the measured
value display with <M>.
Calibration /
Calibration data stor-
age /
Output to RS232/USB
- Outputs the calibration data stor-
age to the interface