EasyManuals Logo

wtw Multi 9310 User Manual

wtw Multi 9310
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
ORP Multi 9310
46 ba75937e03 12/2011
6.1.2 Measuring the temperature
For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
When operating a sensor without integrated temperature sensor, you first have
to measure and enter the temperature of the sample.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
6.2 ORP calibration
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes 0.1 °C Automatic with temperature
sensor
- 1 °C Manual
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 9310 and is the answer not in the manual?

wtw Multi 9310 Specifications

General IconGeneral
Brandwtw
ModelMulti 9310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals